B. Braun Infusomat Space Pump: Easy Setup & Troubleshooting Guide (Singapore)
Share
This is a quick-reference guide to setting up, running and troubleshooting the B. Braun Infusomat® Space volumetric infusion pump. It reproduces the official B. Braun quick guide as fully readable text. Use the search box to jump to a step or an alarm (try "air bubble", "bolus" or "switch off"). Always follow the official B. Braun Instructions for Use and your clinician's guidance.
Steps To Use the Infusomat Space
- Press the on/off button ⏻ and wait for the self-test to complete.
- Press ▼ and open the pump door with OK.
- Feed the primed Infusomat Spaceline into the slot. Insert the infusion line from right to left, keeping it straight — route it first through the upstream sensor, then through the two-hole clip.
- Attach the white clip without twisting the line.
- Insert the free-flow clamp into the opened aperture in the direction of the arrow, until the opening lever locks in and the safety clamp squeezes the line (the flashing signal lamp goes out).
- Close the door.
- Confirm the inserted spaceline with OK.
- Release the roller clamp as instructed by the pump.
- Set the volume to be infused (VTBI) using ▲▼ and confirm with OK.
- Select Rate (ml/hr) or Time, key in the value with ▲▼, and confirm with OK.
- Press ▶‖ Start/Stop to start the infusion.



To Change Rate While Pump is Running
- Press ▶‖ Start/Stop to enter the rate (you do not need to stop the infusion).
- Key in the new rate using the arrow keys ▲▼ and confirm with OK.
To Perform a Bolus with Volume Preselection
- Press the Bolus button (you do not need to stop the infusion).
- Press ▲▼ and set the bolus dose limit (e.g. 5 ml).
- Press the Bolus button to start the bolus. To cancel the bolus at any time, press Bolus again.
Infusion Line Change & New Therapy Start
- Press ▶‖ Start/Stop to stop the infusion and close the roller clamp.
- Press ▼ and open the pump door with OK.
- Press down the green opening lever until it locks in place. Remove the line and insert a new line.
- Close the pump door, confirm the inserted spaceline with OK, and release the roller clamp as instructed.
- In the main menu (infusion stopped), press ≣ Menu.
- Select a new therapy if needed, and set the new parameters with ▲▼.
- Press ▶‖ Start/Stop to start the infusion.
To End an Infusion & Switch Off the Pump
- Press ▶‖ Start/Stop to stop the infusion and close the roller clamp.
- Press ▼ and open the pump door with OK.
- Press down the green opening lever until it locks. Remove the line and close the pump door.
- Press and hold the on/off button ⏻ for 3 seconds to switch the pump off.
Troubleshooting
Air Bubble Alarm
- Press OK.
- Disconnect the connector from the patient.
- Prime 2 ml at a time to expel the bubbles (repeat if necessary). The display shows "Prime line 2 ml?" then "Pump is priming".
- Connect back to the patient.
- Start the infusion.
Check Upstream Alarm
- Check the line from the drip chamber to the pump (roller clamp / swim valve on the burette).
- Press OK.
- Start the infusion.
Pressure High Alarm
- Check the line from the pump to the patient (for kinks or blockages).
- Press OK.
- Start the infusion.
Frequently asked questions
What is the B. Braun Infusomat Space? It is a compact volumetric infusion pump used to deliver fluids and medication at a controlled rate, in hospitals and increasingly in home infusion therapy.
How do I clear an air bubble alarm? Press OK, disconnect the connector from the patient, prime 2 ml at a time to expel the bubbles (repeat if needed), reconnect the patient, then start the infusion.
Can I change the rate without stopping the infusion? Yes. Press Start/Stop to enter the rate, key in the new rate with the arrows, and confirm with OK — the infusion does not need to be stopped.
What is the difference between "Pressure High" and "Check Upstream"? "Check Upstream" points to a problem between the drip chamber and the pump (for example a closed roller clamp or burette valve). "Pressure High" points to a downstream blockage between the pump and the patient (for example a kink or occlusion).
How do I switch the pump off? Stop the infusion, open the door and remove the line, close the door, then press and hold the on/off button for about 3 seconds.
Where can I get Infusomat Space consumables or pump rental in Singapore? EMIS+ supplies B. Braun IV infusion therapy consumables and offers medical appliance rental in Singapore, with fast local delivery and support.